Rockwin have been engaged in designing, developing, producing and calibrating flowmeters for more than two decades. Rockwin's facilities at Sahibabad, near Delhi include a large and sophisticated Calibration Laboratory which is at par with some of the best in the world.
Present calibration facilities
for flowmeters at Rockwin Calibration Laboratory are:
| 500mm Bi-directional Positive Displacement Prover designed and operated to API 2531 | |
| 100mm Uni-directional Positive Displacement Prover designed and operated to API 2531 | |
| Supplementary Transfer Standard loop on 20" Prover | |
| Oil Calibration Rig | |
| Gas Calibration Rig. |
Flowmeters may be calibrated nominally sized from 5 to 400 mm over a flow rate of 0.06 to 3200 M3/Hr subject to individual pressure drop characteristics. Most type of flowmeters may be calibrated provided sufficient signal conditioning equipment is included to provide an acceptable output compatible with the laboratory electronics.
ROCKWIN
CALIBRATION LABORATORY
| Hydraulic Calibration Facility | |
| Oil Calibration Facility | |
| Gas Calibration Facility |
GeneralSpecifications |
|
CALIBRATION RIG |
STANDARD |
MAXIMUM |
CALIBRATION FLUID |
BEST MEASUREMENT |
|
Hydraulic Rig |
500 mm PD Bi-Directional Meter Prover |
1500 |
Water |
+/- 0.016% |
|
Hydraulic Rig |
100mm PD Uni-Directional Meter Prover |
120 |
Water |
+/- 0.012% |
|
Hydraulic Rig |
Transfer Standards |
3200 |
Water |
+/- 0.016% |
|
Gas Rig |
Transfer Standards |
2500 |
Air |
+/- 0.14% |
|
Oil Rig |
Transfer Standards |
18 |
Oil |
+/- 0.02% |